Equipment Reference
A quick reference for day-to-day use of your Oxlife LIBERTY — controls, power options, battery run times, and weekly cleaning. This page is a summary only. Always follow the manufacturer's full user manual and your physician's prescribed settings.
Your prescribed mode and setting come from your physician. Do not change them on your own — call us or your provider if something doesn't look right.
Connect the power supply to the Oxlife LIBERTY, then to the wall outlet. The external power source icon will appear on the control panel.
Connect the DC power cord to the Oxlife LIBERTY, then to the 12/15 volt DC outlet in the vehicle.
Always have the vehicle's engine running before plugging in your Oxlife LIBERTY.
Whenever you have a wall outlet or a vehicle outlet available, use it. Running on external power preserves your battery for the times you actually need it.
The battery and/or device can be damaged if improperly aligned. If it doesn't slide in easily, back it out and re-align it.
| Setting (LPM) | Operating time |
|---|---|
| 0.5 | 2.25 hours |
| 0.75 | 2.25 hours |
| 1.0 | 1.75 hours |
| 1.25 | 1.50 hours |
| 1.5 | 1.25 hours |
| 1.75 * | 1.00 hour |
| 2.0 * | 0.90 hours |
| Setting | Operating time |
|---|---|
| 1 | 3.00 hours |
| 2 | 2.75 hours |
| 3 | 2.50 hours |
| 4 | 2.00 hours |
| 5 | 1.75 hours |
| 6 | 1.50 hours |
| 7 | 1.30 hours |
| 8 | 1.10 hours |
| 9 | 1.10 hours |
| 10 * | 1.00 hour |
* Continuous mode settings 1.75 and 2.0, and pulse mode setting 10, are only available on the Oxlife LIBERTY2® device.
Operating times are based on a new, fully charged battery and are accurate to within ±10%.
Battery operating time will degrade with battery age, number of charge cycles and operating environment.
Clean the side vents once a week using these steps:
Never use your Oxlife LIBERTY without the side vents installed.
Keep a copy in your carry bag or with your travel documents.
This page is a convenience summary prepared by High Country Oxygen from the manufacturer's Quick Reference Guide (document 800-1090 Rev C). It is not a substitute for the complete user manual. For full instructions, warnings and specifications, refer to the manual supplied with your device or visit o2-concepts.com/products. Oxlife LIBERTY®, Oxlife LIBERTY2® and O2 Concepts® are trademarks of their respective owner. Never change your prescribed oxygen settings without direction from your physician.